Pengujian Karakteristik Alat Ukur Radiasi Dose Calibrator Menggunakan Sumber Teknesium-99m di Instalasi Kedokteran Nuklir Rumah Sakit Abdoel Wahab Sjahranie Samarinda

  • Fahira Mutya Mutmainna Program Studi Fisika
  • Pratiwi Sri Wardani Program Studi Fisika, Jurusan Fisika, FMIPA, Universitas Mulawarman
  • Erlinda Ratnasari Putri Program Studi Fisika, Jurusan Fisika, FMIPA, Universitas Mulawarman
  • Retno Zurma Instalasi Kedokteran Nuklir Abdoel Wahad Sjahranie Samarinda

Abstract

Dose Calibrator is one of the dosimeters used to measure radionuclide activity before being given to patients. In order to function as a radiation dosimeter, the dose calibrator must have several good dosimeter properties such as precision and stability. To test the dose calibrator, a source used by the hospital is required. One isotope that is widely used in the field of nuclear medicine, namely Tc-99m. The research steps this time are: firstly the elution performed by radiopharmaceuticals, secondly measuring background activity (background), thirdly calling for Tc-99m sources, fourthly measuring Tc-99m activity, fifthly calculating Tc-99m activity for testing precision, and sixth evaluate the results. Based on the research results on the precision test, it was found that the measurement results for the Tc-99m source were obtained at 0.578%. It can be concluded that the Capintec CRC 25R dose calibrator has good characteristics and performance, so that the measurement results can be trusted.

Downloads

Download data is not yet available.

References

[1] Adler, S., Choyke, P. (2018). Design And Performance Of The Micro-dose Calibrator: Phisics in Medicine & Biology. 3(6). 1-12.
[2] Afismi. (2021). Aliansi Fisikawan Medik Indonesia (AFISMI). https://www.afismi.org/tentang-fisika-medis.
[3] Alatas, Z, Hidayati, S., Akhadi, M., & Purba, M. (2016). Buku Pintar Nuklir. Jakarta: BatanPress.
[4] AntaraNews. (2018). Menkes resmikan instalasi kedokteran miklir RSUD AWS Samarinda. https://www.antaranews.com/berita/696225/menkes-resmikaninstalasikedokteran-nuklir-rsud-aws-samarinda.
[5] Assan, B., Addison, E., Hasford, F., & Sosu, E. (2012). Calibration and Effective Use of a Dose Calibrator: International Journal of Science and Technology. 2(6), 395-400.
[6] Aziz, A. (2013). Evaluasi Kinerja Dose Calibrator Capintec CRC-55 R untuk Pengukuran Aktivitas Radioisotop Yb-175. Prosiding Seminar Nasional Sains Dan Teknologi Nuklir. 4(1), 53-60.
[7] Bapeten. (2012).Jdih bapeten.go.id.
[8] Candra, H. (2016). Pusat Teknologi Keselamatan dan Metrologi Radiasi: Batan. 18(1), 141-465.
[9] Freifelderl, R., Karp, J. S., Geagan, M., & Muehllehner, G. (2018). Design and Performance of the. 41(4), 1436-1440.
[10] IAEA. (2007). Update Of X- And V-Ray Decay Data Standards For Detector Calibration And Other Applications. Autria: IAEA Nuclear Data Section Vienna
[11] Iqlima, M. N. (2020). Kerusakan Sel Hepar Akibat Paparan Radiasi Elektromagnetik Telepon Seluler: Jurnal Kedokteran Dan Kesehatan. 19(1), 40-45.
[12] Karaca, Cinoglu Merve. (2020). Comparative Assessment Of Dose Calibrators Used In Nuclear Medicine: Middle East Journal of Science. 6(2), 44-56.
[13] Mawar, M., Astuti, S. D., & Febriana, S. (2020). Deteksi Karakteristik Peluruhan Radionuklida Medis 99Mo Untuk Memperoleh 99mTc Menggunakan Spektrometer Gamma dengan Sumber Standar Cair Campuran 133Ba dan 152Eu: Fisika Indonesia. 24(2), 96-101.
[14] Mutohar, A., Setiabudi, W., Shintawati, R., Fisika, D., & Diponegoro, U. (2017). Laju Paparan Dan Dosis Radiasi Dari Pasien Terapi Kelainan Kelenjar Tiroid Dengan Pemberian Radiofarmaka lodium-131: Youngster Physics Journal. 6(1), 22-31.
[15] Podgorsak, E.B. (2005). Radiation Oncology Physics: A handbook for teachers and students. Australia: IAEA
[16] Prekeges, J. (2009). Nuclear Medicine Instrumentation In Medical Devices and Systems: Jones & Bartlett Learning.
[17] Purwati, T., Setiabudi, W., & Fisika, J. (2016). Penentuan Waktu Paro Biologi Te 99M Mdp Pada Pemeriksaan Bone Scanning: Youngster Physics Journal. 5(4), 261-268.
[18] Sholikhah, U. N., Lubis, H., Sarmini, E., Herlina, H., & Wisnukaton, K. (2016). Karakteristik Polimer Zirkonium Sebagai Adsorba Generator Mo-99/Tc-99m Untuk Radiofarmaka Diagnostik: Widyariset. 2(1), 17-26
[19] Supriyono, P., Candrawila S., rahim, A. H., & Murni, T. W (2018). Keamanan Peralatan Radiasi Pengion Dikaitkan Dengan Perlindungan Hukum Bagi Tenaga Kesehatan Di Bidang Radiologi Diagnostik: SOEPRA Jurnal Hukum Kesehatan. 3(1), 102-116
Published
2024-06-27
How to Cite
MUTMAINNA, Fahira Mutya et al. Pengujian Karakteristik Alat Ukur Radiasi Dose Calibrator Menggunakan Sumber Teknesium-99m di Instalasi Kedokteran Nuklir Rumah Sakit Abdoel Wahab Sjahranie Samarinda. Progressive Physics Journal, [S.l.], v. 5, n. 1, p. 350-355, june 2024. ISSN 2722-7707. Available at: <https://jurnal.fmipa.unmul.ac.id/index.php/ppj/article/view/1038>. Date accessed: 30 june 2024. doi: https://doi.org/10.30872/ppj.v5i1.1038.